There were multiple medical device types reported to be involved.The information for the additional device type is as follows:
d.2.A medical device type: jka.D.2.B common device name: tubes, vials, systems, serum separators, blood collection.Investigation summary:
bd received 5 photos for investigation.Evaluation of the photos indicated a split female luer adapter.Additionally, 10 retained samples were visually examined, and no cracked luers were found.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has been confirmed for the indicated failure mode: damaged (cracked product/component).Bd initiated further root cause investigation relating to the issue of damaged/cracked luer through corrective and preventive actions.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|