• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Becton Dickinson & CO (Franklin Lakes) BD Vacutainer® UltraTouch¿ Push Button Blood Collection Set; Set, administration, intravascular

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Becton Dickinson & CO (Franklin Lakes) BD Vacutainer® UltraTouch¿ Push Button Blood Collection Set; Set, administration, intravascular Back to Search Results
Catalog Number 367364
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/15/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported while using bd vacutainer® ultratouch¿ push button blood collection set, 1 device had a crack in the female luer resulting in leakage.A repeat venipuncture was required.There was no other health impact or consequences reported.
 
Additional Manufacturer Narrative
There were multiple medical device types reported to be involved.The information for the additional device type is as follows: d.2.A medical device type: jka.D.2.B common device name: tubes, vials, systems, serum separators, blood collection.Investigation summary: bd received 5 photos for investigation.Evaluation of the photos indicated a split female luer adapter.Additionally, 10 retained samples were visually examined, and no cracked luers were found.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has been confirmed for the indicated failure mode: damaged (cracked product/component).Bd initiated further root cause investigation relating to the issue of damaged/cracked luer through corrective and preventive actions.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD Vacutainer® UltraTouch¿ Push Button Blood Collection Set
Common Device Name
Set, administration, intravascular
Manufacturer (Section D)
Becton Dickinson & CO (Franklin Lakes)
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK  
Manufacturer Contact
anand singh
1 becton drive
franklin lakes, NJ 07417
+1(201)847-5186
MDR Report Key50773755
Report Number9617032-2026-01641
Device Sequence Number9488950
Product Code FPA
UDI-Device Identifier30382903673644
UDI-Public(01)30382903673644(11)260127(17)280131(10)6027180
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K212724
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number367364
Device Lot Number6027180
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2026
Initial Report FDA Received Date09/15/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/27/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-