| Model Number |
TCDS0801-XTW |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
08/27/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026 a patient presented with grade 4 functional tricuspid regurgitation (tr), long lateral leaflets and rotated heart for a triclip procedure.During the procedure, while under the valve, leaflets became attached to the shaft and gripper of another clip arm and would not release.There was good grasp of septal leaflet, and the clip was deployed and was stable.Additional clip was attempted but was not implanted due to the anatomy.Imaging was difficult.All steps were performed as per ifu.The final tr was grade 1.There were no adverse patient effects or clinically significant delays reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the lot that would have contributed to the reported event.Additionally, a review of the complaint history identified no other complaints reported from the lot.All available information was investigated, and based on the information provided, a cause for the entrapment of device with the anatomy resulting in foreign body in patient clip cannot be determined.The reported poor visualization appears to be related to procedural circumstances.Unexpected medical intervention was a result of case-specific circumstances.There is no indication of a product issue with respect to manufacture, design or labeling.
|
| |
|