• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Medical TriClip G5 Delivery System; tricuspid valve repair device, percutaneously delivered

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Medical TriClip G5 Delivery System; tricuspid valve repair device, percutaneously delivered Back to Search Results
Model Number TCDS0801-XTW
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 08/27/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) 2026 a patient presented with grade 4 functional tricuspid regurgitation (tr), long lateral leaflets and rotated heart for a triclip procedure.During the procedure, while under the valve, leaflets became attached to the shaft and gripper of another clip arm and would not release.There was good grasp of septal leaflet, and the clip was deployed and was stable.Additional clip was attempted but was not implanted due to the anatomy.Imaging was difficult.All steps were performed as per ifu.The final tr was grade 1.There were no adverse patient effects or clinically significant delays reported.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the lot that would have contributed to the reported event.Additionally, a review of the complaint history identified no other complaints reported from the lot.All available information was investigated, and based on the information provided, a cause for the entrapment of device with the anatomy resulting in foreign body in patient clip cannot be determined.The reported poor visualization appears to be related to procedural circumstances.Unexpected medical intervention was a result of case-specific circumstances.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TriClip G5 Delivery System
Common Device Name
tricuspid valve repair device, percutaneously delivered
Manufacturer (Section D)
Abbott Medical
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, Reg # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
+1(651)756-5400
MDR Report Key50774007
Report Number2135147-2026-06237
Device Sequence Number18513615
Product Code NPS
UDI-Device Identifier05415067050731
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P230007
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberTCDS0801-XTW
Device Catalogue NumberTCDS0801-XTW
Device Lot Number60312R2125
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/27/2026
Initial Report FDA Received Date09/15/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/13/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient SexUnknown
Patient Weight173 LB
-
-