| Model Number |
MT24078 |
| Medical Device Problem Code |
Defective Alarm (1014)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/19/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that an alert/notification settings issue occurred.Product was provided for investigation.An external visual inspection was performed and passed.Receiver charge and boot tests were performed and passed.Functional tests were performed and the vibrator test failed.A manual alarm/alert test was performed and failed.The vibrator motor resistance was measured and failed.An internal visual inspection was performed and passed.The performance log was reviewed finding no errors related to the complaint.The probable cause of no vibration was determined to be a defective vibrator motor.No injury or medical intervention was reported.
|
| |
|
Event or Problem Description
|
|
Verified a defective vibrator motor with high resistance measured ¿0.987 m¿ (known good resistance is 18¿ +/-20%).
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).B5: describe event or problem - correction.H2: correction.
|
| |
|