Customer called in as she had cut her finger and applied the bandage to protect her finger and when she took the bandage off it pulled the skin off as the adhesive was verrry sticky and strong; the customer reported that she did not seek any medical attention, but applied an antibiotic ointment and her finger is now healing.The provided photos were reviewed, the provided samples were evaluated, the vendor was notified of the issue, and trend analysis was performed by reviewing the past two years of complaint data; photos showed a small speck of blood and a used bandage with a small piece of the end user¿s skin still attached to the adhesive backing.Visual inspection of the bandages did not note any defects or possible causes to the reported issue, the bandages were able to adhere to a clean, dry surface and be removed as intended without any signs of aggressive adhesion, and a product defect was not identified from the sample review.Based on the information provided and the results of the investigation, medline can confirm the reported issue took place, however, cannot confirm that the product failed to meet specifications or that the reported issue was related to the manufacture of the product; the vendor¿s review noted no indications of potential causes to the reported issue.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|