| Model Number |
SB-0545FC |
| Medical Device Problem Code |
Peeled/Delaminated (1454)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the instrument for a separate issue.During the initial evaluation of the returned device, it was observed that the tissue pad was peeled off.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the final investigation of the device, the initial evaluation findings were confirmed.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction:
1.The tissue pad was worn away because no tissue was being grasped between the grasping section and the probe tip when the device was activated output in seal & cut mode for/after a transection of tissue.2.The tissue pad was excessively heated due to friction between the grasping section and the probe tip.This caused the tissue pad to be peeled away.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|