BOSTON SCIENTIFIC CORPORATION DYNAGEN? EL ICD VR; Implantable Cardioverter Defibrillator Back to Search Results
Model Number D150
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/04/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the health care professional (hcp) called technical services (ts) for assistance with programming this implantable cardioverter defibrillator (icd) into magnetic resonance imaging (mri) protection mode.Ts confirmed this icd system was mri conditional.Prior to the mri, the shock impedance measurements appeared to be high out of range.The hcp opted to continue with the mri.The mri was successfully completed and the hcp called ts back to confirm the device was out of mri protection mode.Ts confirmed all measurements were within normal limits.No adverse patient effects were reported.This icd remains in service.
 
Additional Manufacturer Narrative
This product investigation is still in progress.This report will be updated when product investigation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN? EL ICD VR
Common Device Name
Implantable Cardioverter Defibrillator
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel E91 T 862
EI   E91 T862
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key50774828
Report Number2124215-2026-49155
Device Sequence Number21438266
Product Code LWS
UDI-Device Identifier00802526534287
UDI-Public(01)00802526534287(17)190408
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P960040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/08/2019
Device Model NumberD150
Device Catalogue NumberD150
Device Lot Number220711
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/04/2026
Initial Report FDA Received Date09/15/2026
Date Device Manufactured05/02/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age49 YR
Patient SexMale
-
-