| Model Number |
N/A |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/21/2026
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
It was reported by the customer that they had a failed io attempt with a defective needle over the weekend.Emts arrived on the scene of a 47 year old male in cardiac arrest.They started high-performance cpr, providing compressions, securing the airway, and attaching the patient to the aed.E207 emts arrived a few minutes later.This crew was tasked with obtaining io access in the right humeral head.The lieutenant removed the io case from their red ems bag and pulled out the drill and yellow needle.He then put the needle onto the drill and located the access point for the patient's right humeral head.He inserted the needle into the arm until he felt it touch the bone and then activated the drill, stopping once the io had been set into place.He pulled the drill straight back and set it next to the patient and then attempted to attach the included tubing only to identify that the internal needle had not been removed from the drill site.He made several attempts to twist the needle to remove it and was unable to get the inner portion free.Three other medics also attempted to remedy the situation with no success.A second io was placed in the tibia and they were able to use this site for the duration of the call.Later, as a paramedic from the original emt crew was working the airway, he reached own, twisted the top of the io to the left, and the io needle pulled right out.Ultimately the patient was pronounced dead at the scene.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Each event reported to bd is evaluated and investigated in accordance with our complaint investigation procedures.The investigation process includes, but is not limited to, evaluation of the event details provided by the complaint facility, a review of complaint history, manufacturing records and risk document where applicable, and an evaluation of the subject device when available to identify potential contributing factors.The device has not been returned to the manufacturer for evaluation.
|
| |
|
Search Alerts/Recalls
|
|
|