| Model Number |
DXW150 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); Visual Disturbances (2140); Halo (2227)
|
| Date of Event |
07/01/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that, following implantation of the preloaded multifocal intraocular lens (iol) in the patient's right eye, a mechanical complication of the ocular implant was identified postoperatively.The lens was subsequently explanted during a secondary surgical procedure and replaced with a different lens model and diopter (den00v 15.5d).Additional information received from the duplicate file confirmed that the replacement lens serial number was (b)(6).The patient experienced visual disturbances with the implanted symfony lens and was noted to have a slight myopic outcome.Postoperatively, the patient also reported halos, glare, and blurry vision.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Section a4, a5 and a6: unknown, information was requested but not provided.Section e1: (email address): unknown/information not provided.Section h6: health effect-impact code:4613- non-serious injury/illness/impairment: (halo, glare and visual disturbance: dysphotopsia: requiring surgical or medical intervention).Section h6: health effect-impact code:4619-temporary impairment: vision blurred (patient daily life activities not impacted).Device evaluation:
product evaluation was not performed because the product has not been received.The complaint issue reported could not be confirmed.Manufacturing record review:
the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search of complaints related to this production order (po) was performed.The search revealed no other complaints were received for this production order number.Conclusion:
based on the investigation, no malfunction or product deficiency was identified.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
|
| |
|