Model Number DXW150
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Blurred Vision (2137); Visual Disturbances (2140); Halo (2227)
Date of Event 07/01/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, following implantation of the preloaded multifocal intraocular lens (iol) in the patient's right eye, a mechanical complication of the ocular implant was identified postoperatively.The lens was subsequently explanted during a secondary surgical procedure and replaced with a different lens model and diopter (den00v 15.5d).Additional information received from the duplicate file confirmed that the replacement lens serial number was (b)(6).The patient experienced visual disturbances with the implanted symfony lens and was noted to have a slight myopic outcome.Postoperatively, the patient also reported halos, glare, and blurry vision.No further information was provided.
 
Additional Manufacturer Narrative
Section a4, a5 and a6: unknown, information was requested but not provided.Section e1: (email address): unknown/information not provided.Section h6: health effect-impact code:4613- non-serious injury/illness/impairment: (halo, glare and visual disturbance: dysphotopsia: requiring surgical or medical intervention).Section h6: health effect-impact code:4619-temporary impairment: vision blurred (patient daily life activities not impacted).Device evaluation: product evaluation was not performed because the product has not been received.The complaint issue reported could not be confirmed.Manufacturing record review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search of complaints related to this production order (po) was performed.The search revealed no other complaints were received for this production order number.Conclusion: based on the investigation, no malfunction or product deficiency was identified.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
Intraocular Lens
Manufacturer (Section D)
AMO Puerto Rico Mfg. Inc.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
+1(714)247-8552
MDR Report Key50775279
Report Number3012236936-2026-000268
Device Sequence Number16105775
Product Code HQL
UDI-Device Identifier05050474762831
UDI-Public(01)05050474762831(17)290426
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberDXW150
Device Catalogue NumberDXW150U130
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/19/2026
Initial Report FDA Received Date09/15/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/26/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexFemale
-
-