MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. HIGH DEFINITION LCD MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. HIGH DEFINITION LCD MONITOR Back to Search Results
Model Number OEV262H
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/26/2026
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported no image during maintenance involving an olympus high definition lcd monitor.There was no patient involvement reported.
 
Additional Manufacturer Narrative
The device was not returned to olympus for inspection, and the reported failure was confirmed via information from technical assistance support (tac).A root cause could not be identified.Based on the results of the investigation, it is likely the most probable cause for no image due to cause traced to component failure.The customer contacted olympus technical assistance center (tac) via phone regarding an image display issue when connecting two oev-262h monitors using a new sdi cable.Troubleshooting steps performed: tac guided the customer to move the sdi connection from sdi 1 to sdi 2 and update the input settings accordingly.Following these steps, the image was displayed on both monitors, and the customer confirmed the issue was resolved.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH DEFINITION LCD MONITOR
Common Device Name
HIGH DEFINITION LCD MONITOR
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50775403
Report Number3002808148-2026-32325
Device Sequence Number9424805
Product Code FET
UDI-Device Identifier04953170363498
UDI-Public04953170363498
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K102379
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 09/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOEV262H
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/26/2026
Initial Report FDA Received Date09/15/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-