Inogen, Inc. Inogen One G5 Portable Oxygen Concentrator Back to Search Results
Model Number IO-500
Medical Device Problem Codes Failure to Charge (1085); Unexpected Shutdown (4019)
Health Effect - Clinical Code Dyspnea (1816)
Date of Event 08/25/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient's device would shut off by itself after making a noise and that the car charger was not working.As a result, the patient experienced difficulty breathing and was hospitalized.No additional information is available at this time as multiple attempts to contact the patient were unsuccessful.
 
Additional Manufacturer Narrative
B3: date of event is estimated.The inogen team attempted to contact the patient for further information three times with no response.The unit came in for no power.The complaint was confirmed with the defective sensor block by water damage that was caused by unit shut down with error 128(b) and data log shows 02 in invalid.Motherboard j 21 and battery board pin damage can have a high impact on the unit (possibly dropped).The data log shows battery low errors; this means patient running unit on low battery.Unit works fine with a good battery.Housing scratch probably user rough cleaning and lower housing broken mount due to compressor vibration.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Inogen One G5 Portable Oxygen Concentrator
Common Device Name
Inogen One G5 Portable Oxygen Concentrator
Manufacturer (Section D)
Inogen, Inc.
600 shiloh rd
plano TX 75074
Manufacturer (Section G)
Inogen, Inc.
600 shiloh rd
plano TX 75074
Manufacturer Contact
greg porter
600 shiloh rd
plano, TX 75074
MDR Report Key50776389
Report Number3004672275-2026-00290
Device Sequence Number10682946
Product Code CAW
UDI-Device Identifier00817131020162
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K032818
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 09/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberIO-500
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/04/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2026
Initial Report FDA Received Date09/15/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/18/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Hospitalization;
Patient SexUnknown
-
-