• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Cordis US. Corp EXOSEAL; Device, Hemostasis, Vascular

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Cordis US. Corp EXOSEAL; Device, Hemostasis, Vascular Back to Search Results
Model Number N/A
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/14/2026
Type of Reportable Event Malfunction
Event or Problem Description
As reported, the deployment button of a 6f exoseal vascular closure device (vcd) could not be pressed.The button depressed only partially and was not able to be fully depressed.Hemostasis was achieved using a suture device.There was no reported patient injury.The exoseal vascular closure device (vcd) and vascular sheath introducer were retracted together until pulsatile flow significantly slowed or stopped from the bleed-back indicator and until the indicator window changed to solid black.The indicator window was solid black at that time and did not show red during retraction.The device was not attempted to be deployed outside the patient's body.A non-cordis 6f short sheath was used.The device was used for an interventional procedure and an antegrade approach was used.Femoral artery suitability, including the vascular sheath introducer insertion angle of 30-45 degrees, was verified by angiography, and the vessel diameter was verified to be at least 5 mm.There was no visible calcium or plaque, access vessel tortuosity, stent near the puncture site, or greater than 50% stenosis at or near the puncture site.The target femoral site had not been previously closed with a closure device or manual compression within 30 days before the procedure.There was no evidence of a pre-existing hematoma, arteriovenous fistula, or pseudoaneurysm at the access site before exoseal vcd use.The patient's body mass index was not greater than 40.The physician had achieved certification for use of the exoseal vcd.The device was stored and prepared according to the instructions for use (ifu).There were no visible signs of device or package damage prior to use.The device will be returned for evaluation.
 
Additional Manufacturer Narrative
Due to system limitations, section h6 required to input type of investigation, investigation findings, and investigation conclusions for an initial report.Pending additional information to complete investigation.This device is available for analysis, but the engineering report is not yet available.However, it will be submitted within 30 days upon receipt.Additional information is pending and will be submitted within 30 days upon receipt.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXOSEAL
Common Device Name
Device, Hemostasis, Vascular
Manufacturer (Section D)
Cordis US. Corp
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer (Section G)
Cordis US Corp.
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer Contact
karla castro
santiago troncoso 808
juarez, chihuahua 32574
MX   32574
+011(786)3138372
MDR Report Key50776626
Report Number9616099-2026-01278
Device Sequence Number10754754
Product Code MGB
UDI-Device Identifier10705032058858
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P100013
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberEX600
Device Lot Number18494567
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/31/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/20/2026
Initial Report FDA Received Date09/15/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/22/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Terumo 6F short sheath.
Patient SexMale
Patient Weight149 LB
-
-