The customer reported a pediatric aneroid that is no longer holding air (bladder).Medline received 1ea hand held aneroid and started testing by attempting to inflate each cuff to 300mmhg; the cuff would not make it past a quarter of inflation, then the bladder was removed from the cuff and the sample was further inspected.The complaint is confirmed, the sample is not working as intended, and a split on the cuff causing a leakage was found.The cause was identified as comp./part defective, broke.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|