H3) the device was sent back to medtronic for evaluation.Upon evaluation of the device, it was determined that the bearing (distal) was misaligned, the bearing (internal tube) was corroded, and the laser markings were illegible/faded.The likely cause could not be determined.H6) annex codes b01, d02, c1701 (misadjustment/misalignment identified), and c0601 (degradation problem identified) are applicable.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|