• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT Powered Surgical Solutions Midas Rex MR8; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT Powered Surgical Solutions Midas Rex MR8; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE Back to Search Results
Model Number MR8-AVS10
Medical Device Problem Code Loss of Power (1475)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/20/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during testing, the device's variable dial was not working/ device seized.There was no patient involvement or impact during testing.
 
Additional Manufacturer Narrative
H3) the device was sent back to medtronic for evaluation.Upon evaluation of the device, it was determined that the bearing (distal) was misaligned, the bearing (internal tube) was corroded, and the laser markings were illegible/faded.The likely cause could not be determined.H6) annex codes b01, d02, c1701 (misadjustment/misalignment identified), and c0601 (degradation problem identified) are applicable.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Midas Rex MR8
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE
Manufacturer (Section D)
MDT Powered Surgical Solutions
4620 n beach st
fort worth TX 76137
Manufacturer (Section G)
MDT Powered Surgical Solutions
4620 n beach st
fort worth TX 76137
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
+1(763)526-5677
MDR Report Key50776714
Report Number1625507-2026-07956
Device Sequence Number16118288
Product Code HBE
UDI-Device Identifier00763000441371
UDI-Public(01)00763000441371(11)251016(20)02
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K183515
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMR8-AVS10
Device Catalogue NumberMR8-AVS10
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/28/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/10/2026
Initial Report FDA Received Date09/15/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/16/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-