• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. EVIS EXERA II ULTRASONIC BRONCHOFIBERVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. EVIS EXERA II ULTRASONIC BRONCHOFIBERVIDEOSCOPE Back to Search Results
Model Number BF-UC180F
Medical Device Problem Codes Material Fragmentation (1261); Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
The customer returned the ultrasonic bronchofibervideoscope to the service center for a separate issue.During the device analysis, the service engineer indicated that deep scratches and a chip in the channel opening were found.There was no patient involvement.
 
Additional Manufacturer Narrative
Date of event is unknown.Additional information will be provided if available.The device was returned to olympus for inspection.Based on the results of the investigation and historical data, reasonably known information suggests probable causes such as damage or deterioration of parts due to wear and tear, without any design or manufacturing issue.The most probable cause of this complaint is expected or random component failure without any design or manufacturing issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II ULTRASONIC BRONCHOFIBERVIDEOSCOPE
Common Device Name
ULTRASONIC BRONCHOFIBERVIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50776900
Report Number3002808148-2026-32358
Device Sequence Number21503513
Product Code PSV
UDI-Device Identifier04953170356360
UDI-Public04953170356360
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K070983
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberBF-UC180F
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/20/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/20/2026
Initial Report FDA Received Date09/15/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/18/2010
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-