| Model Number |
0570-0437 |
| Medical Device Problem Codes |
Power Problem (3010); Intermittent Loss of Power (4016)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer reported that while using a glidescope core 15-inch fhd monitor, the device unexpectedly shut down in the middle of the procedure.No delay in procedure, use of a backup device, or harm to the patient was reported.The end user noted that the battery was fully charged when the shut down occurred.Following the reported event, the facility's biomedical engineering representative was unable to reproduce the reported issue.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported glidescope core 15-inch fhd monitor was returned to verathon for evaluation along with the glidescope core smart cable used during the reported event.A verathon technical service representative evaluated the returned devices but was unable to confirm the reported premature shut down issue.When connecting the monitor to verathon's test equipment, no power issues were observed.The monitor passed visual inspection and verathon's device functionality testing, confirmed to be operating within specification.Upon completion of verathon's device evaluation, both the monitor and smart cable were returned to the customer.No further investigation is required at this time.Verathon will continue to monitor trends.
|
| |
|