Abbott Laboratories Alinity c Processing Module; ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE
|
Back to Search Results |
|
| Catalog Number |
03R67-01 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported elevated magnesium generated from alinity c processing module (sn: (b)(6)) and provided the following data for a 51 year old female patient:
on (b)(6)2026, sample id (b)(4) magnesium result was 6.2 mg/dl.Sample was retested on the analyzer and got a result of 1.9 mg/dl.Sample was also tested twice on another alinity c processing module (sn: (b)(6)) and results were 1.8 mg/dl and 1.9mg/dl.Per the customer discrepant results were not reported out to the medical provider.There was no reported impact to patient management.
|
| |
|
Additional Manufacturer Narrative
|
|
The field service representative (fsr) resolved the issue by replacing multiple components, including the alinity c sample probe.During troubleshooting, gel contamination was observed within the sample probe pathway and on the tip of the sample probe.No additional discrepant result issues have been reported since the service activity was completed.The instrument service history review revealed no additional service ticket associated with discrepant/erratic patient results.There were no additional service or complaint issues on or around the date this complaint was initiated that may have contributed to this issue.A review of complaint and trending data for the alinity c processing module did not identify any similar issues as described in this complaint.Additionally, review of complaint and trending data associated with alinity c sample probe did not identify an increase in complaint activity.A review of the manufacturing documentation did not identify any nonconformances or potential nonconformances.Labeling was reviewed and adequately addresses the customer¿s issue.Based on the investigation, no systemic issue or deficiency of the alinity c processing module for serial number (b)(6)or the alinity c sample probe was identified.All available patient information was included.Additional patient details are not available.
|
| |
|
Search Alerts/Recalls
|
|
|