Roche Diagnostics COAGUCHEK XS PT TEST 6 STRIP; PROTHROMBIN TIME TEST STRIPS Back to Search Results
Catalog Number 04625374160
Medical Device Problem Code Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/21/2026
Type of Reportable Event Malfunction
Event or Problem Description
There was an allegation of questionable inr results on a coaguchek xs pst meter.The patient tested 5 times within 30 minutes.The result at 10:40 a.M.Was 4.5 inr.The result at 10:41 a.M.Was 3.5 inr.The result at 10:43 a.M.Was 3.0 inr.The result at 11:03 a.M.Was 4.2 inr.The result at 11:10 a.M.Was 4.2 inr.The same finger was used for some of the tests, but different fingers were used for the results of 3.0 inr at 10:43 a.M.And 4.2 inr at 11:10 a.M.The patient¿s therapeutic range is 2.0 ¿ 3.0 inr.
 
Additional Manufacturer Narrative
Section e3, occupation is patient/consumer (patient's wife).The coaguchek xs meter serial number was (b)(6).The product was requested for investigation.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK XS PT TEST 6 STRIP
Common Device Name
PROTHROMBIN TIME TEST STRIPS
Manufacturer (Section D)
Roche Diagnostics
9115 hague road
indianapolis IN 46250 -045
Manufacturer (Section G)
Roche Diagnostics GmbH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
natalie catton
9115 hague road
indianapolis, IN 46250
MDR Report Key50777939
Report Number1823260-2026-03769
Device Sequence Number21509861
Product Code GJS
UDI-Device Identifier00365702127104
UDI-Public00365702127104
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K062925
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number04625374160
Device Lot Number89404424
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/21/2026
Initial Report FDA Received Date09/15/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexMale
-
-