Customer (b)(6) reported on 10aug2026 discrepant hbv results with a donor sample that tested nonrepeat reactive on the ultrio elite screening assay on (b)(6) 2026 and (b)(6) 2026.A fresh sample was initially taken from the donor on (b)(6) 2026 that tested reactive on ultrio elite (3x), hbv discriminatory and serology (anti-hbs and anti-hbc).Hbsag, hbe ag and anti- hbe were negative.A look back at the donor results showed that a sample drawn from the same donor on 12dec2025 was negative for ultrio elite and serology (hbs ag).The archived sample was retested on (b)(6) 2026 with non repeat reactive results for ultrio elite, nonreactive hbv discriminatory results and positive results for anti-hbc (negative anti-hbs).A fresh draw from the donor was taken on (b)(6) 2026 and tested on ultrio elite in duplicate resulting in nonrepeat reactive results.The first sample tested positive at 17 iu/ml on qhbv pcr (abbott alinity) but negative with the second sample tested.The status of the blood donation drawn on (b)(6) 2025 and (b)(6) 2026 is pending.Summary of customer testing:
(b)(6) 2025: initial fresh draw
ultrio elite ml 710006
s/co, nonreactive
serology (abbott alinity)
hbsag nonreactive.(b)(6) 2026: fresh draw
ultrio elite ml 710333
s/co, reactive (3x)
hbv discriminatory, reactive
serology (abbott alinity)
hbsag nonreactive
anit-hbs (10 iu/ml) reactive
anti-hbc (6.5 s/co) reactive
hbe ag nonreactive
anti-hbe nonreactive.(b)(6) 2026: archived sample from (b)(6) 2025
ultrio elite ml 710333
s/co, reactive
s/co, nonreactive
hbv discriminatory, nonreactive.Serology (abbott alinity)
hbsag nonreactive
anit-hbs (8.28 iu/ml) nonreactive
anti-hbc (6.29 s/co) reactive
hbe ag nonreactive
anti-hbe nonreactive.(b)(6) 2025: fresh draw
ultrio elite ml 710333
sample 1:
s/co, reactive
qhbv pcr (abbott alinity)
17 iu/ml, reactive.Sample 2:
s/co, reactive
s/co, nonreactive (2x)
qhbv pcr (abbott alinity)
nonreactive.An investigation is ongoing for the discrepant hbv results.Remaining sample from the donor is pending availability for investigational testing.No related quality events were found for the lots/batches of product and the products met all release criteria.A previous events search did not indicate a trend for hbv sensitivity issues with procleix ultrio elite ml 710006 or ml 710333.
|