It was reported that device shift occurred.A left atrial appendage (laa) closure procedure was performed, and a 35mm watchman flx pro closure device was implanted on (b)(6) 2026.During the procedure there were no complications, there was only one (1) partial recapture, and the closure device met the position, anchor, size and seal (pass) criteria and was released.The patient was discharged.On (b)(6) 2026, the patient had been admitted for blurry vision, and upon further examination a transesophageal echocardiography (tee) was performed, and the location of the closure device was noted as proximal when compared to the implant images.A computed tomography (ct) scan was requested by the physician.Once the scan is assessed the physician will determine if it needs to be explanted or not.
|