Catalog Number DYNJ03108LF
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/17/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "as soon as the provider tried to suture with the forceps the suture needle kept falling out." it was also reported that "it looks like a crack going through" and "when they close the ratchet the jaws do not always close completely.".
 
Additional Manufacturer Narrative
The customer reported that during patient use, the provider tried to suture with the forceps and the suture needle kept falling out; the customer also reported that it looked like a crack going through and that when providers close the ratchet the jaws do not always close completely.A photo and a sample were received for investigation, the sample was evaluated, finished good production records were reviewed, and the finished good pack build was reviewed.One of the needle holders received was observed to have a crack in the jaw, confirming the reported issue.Because medline does not alter the component before placing it in the kit, the root cause was determined to be a vendor/supplier product issue relating to the component manufacturing process; the complaint was provided to the component manufacturing facility for further evaluation and the supplier quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
LACERATION TRAY
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
nitish sandadi
three lakes drive
northfield, IL 60093--275
MDR Report Key50778307
Report Number1417592-2026-41316
Device Sequence Number9762138
Product Code MCY
UDI-Device Identifier10080196841202
UDI-Public10080196841202
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYNJ03108LF
Device Lot Number24LBU320
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer05/09/2025
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/17/2025
Initial Report FDA Received Date09/15/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-