The reported product problem/issue was not holding air; the complaint transcript states the pad "didn't hold air after putting it on the bed." the customer sent in only the pad without the pump, and medline was unable to evaluate without the pump to inflate the pad.The complaint was unable to be confirmed/not confirmed with received sample, and no root cause was identified.There was no reported death or serious injury, product was not being used for treatment or diagnosis, and the complaint trending report showed no negative trends for this defect or product code, indicating no need for further action at this time.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|