|
The account reported finding coban adhesive issue, and the complaint text stated that the coban is very sticky and hard to unwrap.No sample or photo was received for the investigation; finished good production records were unable to be reviewed due to no lot number being provided, and the finished good pack build was reviewed with no issues noted with the interior placement of the components that could contribute to the issue reported.At this time, medline was unable to confirm the reported defect, as no physical sample or detailed photographic evidence was provided to support the investigation; the cause was sample required to determine cause.The division will monitor this issue for trending purposes, and there has been no change to the reportability decision associated with this complaint.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|