It was reported, spinal tray, failed spinal, reverted to general anesthesia.Patient 2 of 3.The following was provided by the initial reporter: i am reaching out as our or team advised yesterday that they reported 4 failed spinals in the past couple days due to the marcaine in these 405672 trays not working.Trays were ordered under purchase order (b)(4) and lot of the trays was reported to be b01w139d with the lot of the faulty marcaine ((b)(6)) in those trays being lp8046.I did try to file a case on the customer care portal however that kept giving me an error when trying to submit without telling me what the error was.Please let us know if there is any other information needed in order to investigate this.Additional information:
can you please confirm an exact date of event in format dd-mmm-yyyy? 19-aug-2026 and 18-aug-2026.Can you please provide the lot number associated with reported issue? item 405672 has a lot b01w139d, marcaine contained in the kit (hospira (b)(6)) that was reported to have failed has a lot lp8046.Any sample or photo available for investigation: no.Describe any patient harm, injury, complication or negative outcome that occurred as a result of the event: 4 failed spinals in 2 days due to the marcaine not working.Could you please provide a further description of the issue? bd spinal tray ref 405672 that carries marcaine 2percent for the injection contained marcaine that did not work on patients when used resulting in or team reporting 4 failed spinals within two days.Did this occur with four different patients? yes, this occurred with 4 different patients.Did they have to convert to general anesthesia or was the procedure completed with a new vial of medication? 3 out of the 4 were converted to general anesthesia with the last one working initially but wore off within an hour or hour and a half of placement.Team cannot remember if doctor said if he converted to general or not on the one that initially worked.
|