| Model Number |
IO-500 |
| Medical Device Problem Codes |
Failure to Charge (1085); Failure to Power Up (1476); Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
Dyspnea (1816)
|
| Date of Event |
08/25/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient's device would shut off by itself and would not turn on.It was noted that the car charger would not work as well.The patient did not have a backup during the event.As a result, the patient experienced shortness of breath until the replacement was received.The patient has no issues with the replacement device.
|
| |
|
Additional Manufacturer Narrative
|
|
B3: date of event is estimated.The unit came in for power port damaged.The complaint was confirmed with the mother board (j14) pin burn & damaged power input.This can happen with the overcurrent due to low voltage events.The data log shows numerous battery low errors.This means the patient runs the unit on low battery and o2 invalid that caused muffler to fill with dust.
|
| |
|