Model Number IO-500
Medical Device Problem Codes Failure to Charge (1085); Failure to Power Up (1476); Unexpected Shutdown (4019)
Health Effect - Clinical Code Dyspnea (1816)
Date of Event 08/25/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient's device would shut off by itself and would not turn on.It was noted that the car charger would not work as well.The patient did not have a backup during the event.As a result, the patient experienced shortness of breath until the replacement was received.The patient has no issues with the replacement device.
 
Additional Manufacturer Narrative
B3: date of event is estimated.The unit came in for power port damaged.The complaint was confirmed with the mother board (j14) pin burn & damaged power input.This can happen with the overcurrent due to low voltage events.The data log shows numerous battery low errors.This means the patient runs the unit on low battery and o2 invalid that caused muffler to fill with dust.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Inogen One G5 Portable Oxygen Concentrator
Common Device Name
Inogen One G5 Portable Oxygen Concentrator
Manufacturer (Section D)
Inogen, Inc.
600 shiloh rd
plano TX 75074
Manufacturer (Section G)
Inogen, Inc.
600 shiloh rd
plano TX 75074
Manufacturer Contact
greg porter
600 shiloh rd
plano, TX 75074
MDR Report Key50780232
Report Number3004672275-2026-00291
Device Sequence Number16141056
Product Code CAW
UDI-Device Identifier00817131020162
UDI-Public(01)00817131020162
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K032818
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 09/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberIO-500
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/08/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/26/2026
Initial Report FDA Received Date09/15/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/19/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;