Model Number CH-S400-XZ-EB
Medical Device Problem Codes Leak/Splash (1354); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/20/2026
Type of Reportable Event Malfunction
Event or Problem Description
The customer returned the camera head for a separate issue.During the initial evaluation of the returned device, a loss of watertightness and the coupler was found to detach from the scope were observed.No patient involvement.
 
Additional Manufacturer Narrative
The final investigation has been completed.The failure identified during the inspection have been confirmed a definitive root cause could not be identified.Based on the results of the investigation, the most probable cause of the complaint is cause traced to component failure.For the coupler detaches from the scope due to deterioration of the coupler unit, water tightness is lost due to the focus ring was cracked.Olympus will continue to monitor the field performance of this device and evaluate any additional information received through post-market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4K CAMERA HEAD
Common Device Name
CAMERA HEAD
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50782601
Report Number3002808148-2026-32458
Device Sequence Number10516069
Product Code FET
UDI-Device Identifier04953170374609
UDI-Public04953170374609
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
K172817
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 09/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCH-S400-XZ-EB
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/02/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/20/2026
Initial Report FDA Received Date09/16/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-