• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OSSIO Ltd. OSSIOfiber® Compression Staple System; ORTHOPEDIC TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OSSIO Ltd. OSSIOfiber® Compression Staple System; ORTHOPEDIC TRAY Back to Search Results
Model Number OF2061110S
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/18/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Revision about 1 year following a 4-corner fusion procedure.
 
Additional Manufacturer Narrative
Company was informed of a revision procedure of 2 staples.An investigation was performed including internal analysis of records.No imaging was provided.No information on clinical symptoms was shared.Revision surgery was performed successfully.Fragments of the removed implants were returned to the company and examined.There is no indication of a design, manufacturing or process issue affecting implant safety, effectiveness or sterility.Based on the limited information available the root cause leading to revision surgery could not be determined.Several known inherent risks, both procedure and patient related may lead to additional surgeries and device removals.As a revision surgery was performed and because the company cannot rule out the possible contribution of the device to the event, out of an abundance of caution, this event is being reported.Company continues to monitor these events as part of post market activities.Additional reports relating to this event are: # 3014323288-2026-00009 and 3014554088-2026-00008/9.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSIOfiber® Compression Staple System
Common Device Name
ORTHOPEDIC TRAY
Manufacturer (Section D)
OSSIO Ltd.
8 hatochen st.
caesarea, 30798 61
IS  3079861
MDR Report Key50782803
Report Number3014323288-2026-00008
Device Sequence Number9570825
Product Code OJH
UDI-Device Identifier07290019361039
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/16/2026
Report Date (Section F) 09/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberOF2061110S
Device Catalogue NumberOF2061110S
Device Lot NumberOF05101
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/24/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date08/19/2026
Device Age9 MO
Event Location Hospital
Date Report to Manufacturer08/19/2026
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date09/16/2026
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OSSIOfiber® Compression Staple
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexFemale
Date Report Sent to FDA09/16/2026
-
-