Centinel Spine, LLC Prodisc C Vivo US Implant, Extra Large, 5mm; PROSTHESIS, INTERVERTEBRAL DISC
|
Back to Search Results |
|
| Model Number |
PDVXL5 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/09/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A patient was implanted with a 2-level prodisc c vivo device below a fusion at c5-7 on (b)(6) 2026.There was no report of patient pain or other symptoms.However, imaging showed that the implant at c5-6 had migrated anteriorly.The implant was removed from c5-6 on (b)(6) 2026 and replaced with a prodisc c nova device.
|
| |
|
Additional Manufacturer Narrative
|
|
An mdr is indicated for this complaint.A patient was implanted with a 2-level prodisc c vivo device below a fusion at c5-7 on (b)(6) 2026.There was no report of patient pain or other symptoms.However, imaging showed that the implant at c5-6 had migrated anteriorly.The implant was removed from c5-6 on (b)(6) 2026 and replaced with a prodisc c nova device.A dhr review found no anomalies associated with the complaint.Complaint trending found that the rate of complaints is within the level outlined in the risk documentation.A review of the risk documentation found that the hazards associated with the complaint are identified and mitigated to a level where the clinical benefits outweigh the harms.The implant was not returned following the surgery; however, imaging was provided that showed that the implant had migrated anteriorly.There were no anomalies identified during the complaint investigation.The removal was completed due to implant migration.The submission is 1 of 1 device involved in this event.
|
| |
|
Search Alerts/Recalls
|
|
|