• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. Bard, Inc. (BASD) -3006260740 Navarre Opti Drain; Chronic Catheters

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. Bard, Inc. (BASD) -3006260740 Navarre Opti Drain; Chronic Catheters Back to Search Results
Catalog Number NOD10LPT
Medical Device Problem Codes Break (1069); Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/27/2026
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6)2026, a patient underwent a ultrasound guided percutaneous percutaneous drainage catheter placement procedure using navarre opti drain drainage catheter.During the procedure, the sure loktm thread was allegedly broken.The affected drainage catheter was replaced with a new drainage catheter, and the procedure was completed successfully.There was no reported patient injury.
 
Event or Problem Description
On (b)(6) 2026, a patient underwent a ultrasound guided percutaneous percutaneous drainage catheter placement procedure using navarre opti drain drainage catheter.During the procedure, the sure loktm thread was allegedly broken.The affected drainage catheter was replaced with a new drainage catheter, and the procedure was completed successfully.There was no reported patient injury.
 
Additional Manufacturer Narrative
H11: as the lot number for the device was provided, a review of the device history records will be performed.The device has not been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.Section a through f: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
H11: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this lot.Investigation summary: although the physical device was not returned for evaluation, two photos were provided for review.The first photo shows the partial view of a clinician holding the drainage catheter and thread.The thread noted to be completely detached from the proximal hub thread hole.The second photo shows the complete view of a fully detached thread and the drainage catheter is not visible in the provided photo.Therefore, the investigation is confirmed for the reported thread break and material separation issues for both the devices.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.G3, h6 (method, result, conclusion).Section a through f : the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Navarre Opti Drain
Common Device Name
Chronic Catheters
Manufacturer (Section D)
C.R. Bard, Inc. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
C.R. Bard, Inc. (GFO)
289 bay road
queensbury 12804
Manufacturer Contact
michael gaffney
800 w. rio salado pkwy
tempe, AZ 85281
+1(480)303-2689
MDR Report Key50784559
Report Number3006260740-2026-06766
Device Sequence Number21213585
Product Code GBX
UDI-Device Identifier00801741037030
UDI-Public(01)00801741037030(17)280728(10)GFKT2336
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
Unknown
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberNOD10LPT
Device Lot NumberGFKT2336
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/27/2026
Supplement Date Received by Manufacturer09/21/2026
Initial Report FDA Received Date09/16/2026
Supplement Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-