| Model Number |
0684-00-0568-01U |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
08/26/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
During an emergency support program call, a cvicu rn requested assistance on behalf of a physician regarding repositioning of a sensation plus 7.5fr.40cc intra-aortic balloon (iab) catheter.The physician reported that the chest x-ray showed the radiopaque distal marker approximately 6 cm below the carina.Further investigation determined that the patient had experienced a cardiac arrest, during which the iab catheter migrated to approximately the 6th-7th intercostal spaces, followed by a decrease in arterial pressure readings.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon completion of the investigation into this event a follow up report will be submitted.This event contains both an adverse event and a product problem.Due to a temporary system limitation associated with recent fda emdr system enhancements, only the adverse event designation could be transmitted in the current submission.Product problem details are included within this report.A supplemental mdr will be submitted once the emdr utility has been updated to support transmission of both event classifications.
|
| |
|