Model Number 0998-00-0800-52
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/27/2026
Type of Reportable Event Malfunction
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Updated field: b1, b4,d10, g3, g6, g7,h2, h6(type of investigation, investigation findings, investigation conclusions), h11 corrected field: e2, g2 kindly revert a1(patient id) and h4 as it should be kept blank.Due to characters limiti in e1 event site address is (b)(6) a getinge field service engineer (fse) evaluated the device and provided loaner batteries to replace the existing batteries.The removed batteries were returned to the service office for the cardiosave battery calibration process.One calibrated battery set was returned and installed in the device.Following installation, the unit was tested and operated as intended.The device successfully passed all performance verification tests in accordance with factory specifications.The device was returned to the customer and cleared for clinical use.
 
Additional Manufacturer Narrative
Event site name: (b)(6).Upon completion of the investigation into this event a follow up report will be submitted.
 
Event or Problem Description
It was reported by the end user that during routine check, the cardiosave intra-aortic balloon pump (iabp) displays battery maintenance due message.There was no patient involvement.No harm is reported.A getinge field service engineer (fse) was dispatched to evaluate the unit.The fse reported that batteries calibration process was performed.Product passed all functional and safety tests per manufacturer specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE G PLUG
Common Device Name
System, Balloon, Intra-Aortic And Control
Manufacturer (Section D)
Datascope Corp.
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
Datascope Corp.
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
ewelina chorlowska
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key50784835
Report Number2249723-2026-0004304
Device Sequence Number18523018
Product Code DSP
UDI-Device Identifier10607567108438
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K112372
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 09/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0998-00-0800-52
Device Catalogue Number0998-00-0800-52
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/27/2026
Initial Report FDA Received Date09/16/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/28/2022
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-