Model Number B35300
Medical Device Problem Code Communication or Transmission Problem (2896)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that information was received from a patient regarding an external device.The reason for call was that the caller reports that they are having trouble with the recharger.The caller indicates that the connection keeps breaking between the recharger and ins all the time and it takes forever to charge the implanted neurostimulator.The issue has been going on for "quite a while" and it seems to be getting worse.They have not had any falls, trauma, or significant changes to weight.Helped caller use the recharger without the drape and after repositions they were able to see excellent connection and go from 90% to 100% while on the call.Caller noted that the harness is too big for them and it feels like it is strangling them.A smaller drape was sent.The caller will continue to monitor recharging.The caller was not able to find the serial number of the recharger on the handset and agent did not want them to pull it away to get the serial number off the back.Additional update was received from patient stating the replacement device was better but still it was too long and they just held it in place while being laid down.
 
Additional Manufacturer Narrative
Continuation of d10: product id: fp6000s, lot# serial#: unknown, implanted: on (b)(6) 2025, product id: neu_rechar ger_acc, lot# serial#: unknown, product type: recharger.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Percept
Common Device Name
STIMULATOR, ELECTRICAL, IMPLANTED, FOR ESSENTI
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
+1(763)526-5677
MDR Report Key50785098
Report Number3004209178-2026-15636
Device Sequence Number18523020
Product Code PJS
UDI-Device Identifier00763000564346
UDI-Public(01)00763000564346(17)270528
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P960009
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberB35300
Device Catalogue NumberB35300
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/26/2026
Initial Report FDA Received Date09/16/2026
Date Device Manufactured06/04/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"See H11."
Patient SexFemale
-
-