A device history record, complaint history review, and risk review on failure modes are performed on each device with reportable malfunction(s) along with other methods of investigation as coded in section h6 of this mdr report.Investigation summary:
field technician stated issue of plunger head - damaged/stuck was confirmed.¿ on (b)(6) 2026, pca was received unboxed and with paperwork.¿ passed asm binary sensors test.¿ carriage block mis-assembled.¿ workmanship at bd manufacturing.¿ device to be stored for future processing.¿ noted issue(s) were corrected in bd fio lab.No repair required.¿ failure investigation report attached to sales force.This report lists imdrf annex a, c, d, and g codes that are reportable malfunctions observed on the device during servicing
per 803.52(f)(11)(iii) the information provided was obtained from servicing activities performed on the device.There were no additional details obtainable or provided at the time of service.
|