| Model Number |
SC-4318 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Codes |
Fall (1848); Muscle Weakness/Atrophy (1967); Pain (1994); Seroma (2069); Discomfort (2330); Numbness (2415)
|
| Date of Event |
07/08/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The patient experienced discomfort at the implant site, described as sharp and aching pain occurring daily along with generalized weakness in the legs.The patient also reported being unable to wear a belt because it contacted the painful area.The patient continued taking pregabalin, which was reported to be helpful.Imaging confirmed that the leads had migrated causing discomfort near the anchor site.The patient underwent a lead replacement procedure.The patient was reported to be doing well postoperatively.The facility kept the explanted product.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional suspect medical device component involved in the event:
model number/catalog number: sc-2318-50
serial number: (b)(6)
batch/lot number: 5005669
model/catalog description: infinion pro lead kit, 16 contact, 50cm
unique identifier (udi) #: (b)(4)
model number/catalog number: sc-2318-50
serial number: (b)(6)
batch/lot number: 5005790
model/catalog description: infinion pro lead kit, 16 contact, 50cm
unique identifier (udi) #: (b)(4).
|
| |
|