| Model Number |
E160 |
| Medical Device Problem Codes |
Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Code |
Shock from Patient Lead(s) (3162)
|
| Date of Event |
09/03/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the health care professional (hcp) requested a review of reports for this implantable cardioverter defibrillator (icd).The hcp suspected this icd delivered anti-tachycardia pacing (atp) and shocks for what appeared to be supraventricular tachycardia (svt) with rapid ventricular response (rvr).Technical services (ts) reviewed the data and noted the rhythm did appear to be svt with rvr and recommended the patient follow up with cardiology.No additional adverse patient effects were reported.This icd remains in service.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|