It was reported that during an unspecified arthroscopic surgical procedure, while using an unknown three-pole electrode device, the surgeon manually adjusted the power of the equipment to its maximum level, which resulted in water spraying at a very high temperature, directly impacting the patient's operated arm.It was further reported that the hot fluid caused a superficial burn on the first layer of the skin, accompanied by the formation of blisters in the affected area.It was reported that, at the end of the surgery, the injured area would be covered using a special dressing for burns.It was reported that the patient was stable with burns.All available information has been disclosed.
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative:
since the complaint device was discarded, we cannot determine a root cause for the reported failure.If the device or additional information is received in the future, we will reopen the complaint and perform the investigation as appropriate.It has been determined that no corrective or preventive actions are proposed.There is no indication that a design or manufacturing issue caused the reported condition.As part of depuy synthes quality processes, all devices are manufactured, inspected, and released in accordance with approved specifications.Additional monitoring for potential safety signals is conducted through complaint trend analysis as part of post-market safety surveillance activities.Device history review: given that no part number was provided, a manufacturing record evaluation (mre) review cannot be performed.If the part number becomes available, the mre review will be performed.
|