• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OGDEN MANUFACTURING PLANT OPTIFLUX F160NRe HF 12/CS 1.5SA EBEAM; Dialyzer, High Permeability With Or Without Sealed Dialysate System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OGDEN MANUFACTURING PLANT OPTIFLUX F160NRe HF 12/CS 1.5SA EBEAM; Dialyzer, High Permeability With Or Without Sealed Dialysate System Back to Search Results
Catalog Number 0500316E
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 09/08/2026
Type of Reportable Event Malfunction
Event or Problem Description
A user facility¿s clinic manager (cm) reported that a visible internal dialyzer blood leak occurred immediately after the initiation of the patient¿s hemodialysis (hd) treatment.The machine, a fme 2008h machine, alarmed appropriately with a blood leak alarm.Blood leak test strips were not used because the leak was visible.A fme bloodline was in use.There was no obvious defect or damage seen on the dialyzer.The patient¿s estimated blood loss (ebl) was approximately 125ml.The patient was restarted on the same machine and treatment completed successfully with new supplies.There was no patient injury or medical intervention required due to the reported event.The complaint device is not available to be returned to the manufacturer for evaluation.One photo has been provided.
 
Additional Manufacturer Narrative
Plant investigation: as the device was not returned to the manufacturer, a physical evaluation could not be performed.However, one photograph was provided by the customer.The photograph showed the dialyzer label from the reported lot number.No damage, defect, or leak was visible in the photograph.A specific cause of the leak was not visible in the photograph.A production records review was performed on the reported lot.An investigation of the device history records (dhr) was conducted by the manufacturer.There was one non-conformance (nc) in the production of this lot.It is unrelated to the reported complaint event.There was no indication of product nonacceptance, deviation, non-conformance, rework, labeling or process control failure during the manufacturing process which could be associated with the reported event.The lot met all release criteria.The investigation into the complaint was not able to confirm the reported event.Continuous improvement is of the utmost importance to fresenius medical care as we strive to provide dialysis products of the highest quality to our patients.Reports of leaking product are investigated both individually as complaints, as well as via the nc/capa program, in order to assess and improve our products and processes.Capas for vision systems and blood leak reduction are recent examples of leak related investigations directed at an overall reduction in dialyzer leaks.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLUX F160NRe HF 12/CS 1.5SA EBEAM
Common Device Name
Dialyzer, High Permeability With Or Without Sealed Dialysate System
Manufacturer (Section D)
OGDEN MANUFACTURING PLANT
director, site quality
475 west 13th street
ogden UT 84404
Manufacturer (Section G)
OGDEN MANUFACTURING PLANT
director, site quality
475 west 13th street
ogden UT 84404
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
+1(617)417-5172
MDR Report Key50787017
Report Number0001713747-2026-00345
Device Sequence Number9415895
Product Code KDI
UDI-Device Identifier00840861100149
UDI-Public00840861100149
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K162488
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0500316E
Device Lot Number26DU06028
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2026
Supplement Date Received by Manufacturer09/26/2026
Initial Report FDA Received Date09/16/2026
Supplement Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/24/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FME Bloodlines; FME Hemodialysis Machine
Patient SexFemale
Patient Weight138 LB
-
-