• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Cardinal Health 200, LLC Kangaroo; Pump, Infusion, Enteral

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Cardinal Health 200, LLC Kangaroo; Pump, Infusion, Enteral Back to Search Results
Model Number BSPFD
Medical Device Problem Code Air/Gas in Device (4062)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/24/2026
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported the pump did not stop at the end of the feed and after the bag was empty it created air in the tube.The customer admitted he had the pump for one year and never cleaned it or the sensors.The customer was educated on cleaning the pump and sensors.
 
Additional Manufacturer Narrative
The product id and lot number are unknown.An investigation is currently underway.Upon completion, the results will be forwarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Kangaroo
Common Device Name
Pump, Infusion, Enteral
Manufacturer (Section D)
Cardinal Health 200, LLC
3651 birchwood drive
waukegan IL 60085
Manufacturer (Section G)
Cardinal Health
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
3651 birchwood drive
waukegan, IL 60085
+1(508)618-3640
MDR Report Key50787505
Report Number1423537-2026-00291
Device Sequence Number15813093
Product Code LZH
UDI-Device Identifier10192253068900
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K221603
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Type of Report Initial,Followup
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberBSPFD
Device Catalogue NumberBSPFD
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/02/2026
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date09/16/2026
Supplement Report FDA Received Date09/23/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Patient Sequence Number1
Patient SexUnknown
-
-