Information contained in this report was previously submitted through asr on 19/oct/2015.A review of the device history record has been completed.No deviations or non-conformances noted.The event of "capsular contracture" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation and capsular contracture, baker grade iv
device evaluation:
per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:
¿ capsular contracture: unable to observe through visual inspection as it is a physiological phenomenon.¿ deflation: observed an opening through microscopic inspection assessed as fold flaw opening.No further actions are required as it is not related to the manufacturing process.None of the other observations performed during the device analysis wear abrasion and creases are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|