(b)(4).Date sent 9/28/2026.D4 batch # z7050n.Investigation summary:
the product was returned to lab cincinnati for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the er320 device was received with the jaws closed and no apparent external damage.The device underwent ct scanning to evaluate its internal components, the trigger was found bent.To evaluate the reported incident, the device was subjected to functional testing.During testing, the device couldn¿t be fired as the trigger is bent.Furthermore, twenty-one (21) clips were found remaining within the clip track.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached as to what may have caused the reported incident.A manufacturing record evaluation was performed for the finished device batch z7050n number, and no non-conformances were identified.
|