• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Biotronik SE & Co. KG Pamira S DX 65/15; ICD lead

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Biotronik SE & Co. KG Pamira S DX 65/15; ICD lead Back to Search Results
Model Number 428775
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 09/09/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Combination product: yes.An infection was observed following the implantation of this biotronik device.The sterilization process was investigated.The validated process assures that all sterilization parameters, such as gas concentration, temperature, humidity, etc., are within its specified ranges for each distributed device.Additionally, an analysis of validated microbiological indicators is performed after every sterilization procedure as evidence of successful completion of the sterilization process.Review of the biotronik complaint database did not reveal any changes regarding the trend for this type of incident.In summary, the infection was not device related.
 
Event or Problem Description
System was explanted due to infection.Should additional information be received, this file will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Pamira S DX 65/15
Common Device Name
ICD lead
Manufacturer (Section D)
Biotronik SE & Co. KG
woermannkehre 1
berlin 12359
GM  12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
+1(877)245-9800
MDR Report Key50788582
Report Number1028232-2026-04292
Device Sequence Number18769320
Product Code NVY
UDI-Device Identifier04035479156510
UDI-Public(01)04035479156510(17)280430
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P980023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number428775
Device Catalogue Number428775
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2026
Initial Report FDA Received Date09/16/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/27/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
-
-