| Model Number |
SC-1110-02 |
| Medical Device Problem Codes |
Nonstandard Device (1420); Adverse Event Without Identified Device or Use Problem (2993); Wireless Communication Problem (3283)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/20/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient has difficulty connecting remote control to scs.It was noted that the patients implantable pulse generator (ipg) was no longer functioning as intended.The patient underwent a spinal cord stimulation (scs) system replacement procedure.During the revision the physician noticed fraying of the paddle lead.The patient was doing well postoperatively.The explanted device components were kept by the facility.
|
| |
|
Additional Manufacturer Narrative
|
|
Block b3: exact date unknown, explant date used as the approximated date of event.Block d2b: qrb
additional suspect medical device components involved in the event:
model number/catalog number: sc-8120-70
serial number: (b)(6)
batch/lot number: 227336a
model/catalog description: artisan surgical lead 70cm
unique identifier (udi) #: (b)(4).
|
| |
|