| Catalog Number |
0500318E |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
09/08/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A user facility¿s registered nurse (rn) reported that an external dialyzer blood leak occurred with one hour remaining in a patient¿s hemodialysis (hd) treatment.The patient¿s estimated blood loss (ebl) is 10ml.The leak originated from the arterial port of the dialyzer.The complaint device is not available to be returned to the manufacturer for evaluation.Two photos have been provided.Multiple attempts have been made to contact the customer to obtain additional information related to the reported event.At this time, no additional information has been provided.
|
| |
|
Event or Problem Description
|
|
A user facility¿s registered nurse (rn) reported that an external dialyzer blood leak occurred with one hour remaining in a patient¿s hemodialysis (hd) treatment.The machine, a fme 2008t machine, did not alarm but was not expected to.A fme bloodline and dialyzer were in use.The leak originated from the arterial connection between the bloodline and dialyzer.It is unclear if the cause of the blood leak was a dialyzer or bloodline malfunction.The patient¿s estimated blood loss (ebl) was approximately 10ml.The patient did not continue treatment after the incident because they only had 45 minutes remaining.There was no patient injury or medical intervention required due to the reported event.The complaint device is not available to be returned to the manufacturer for evaluation.Two photos have been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: as the device was not returned to the manufacturer, a physical evaluation could not be performed.However, two photographs were provided by the customer.Both photographs showed the dialyzer connected to the machine (from different views) and the arterial bloodline attached to the dialyzer blood port.There was blood leaking from the connection between the bloodline and the dialyzer blood port onto the header cap and down the side of the housing external to the dialyzer.There was no visible damage on the blood port.A specific cause of the leak was not visible in the photograph.A production records review was performed on the reported lot.An investigation of the device history records (dhr) was conducted by the manufacturer.There was no indication of product nonacceptance, deviation, non-conformance, rework, labeling or process control failure during the manufacturing process which could be associated with the reported event.The lot met all release criteria.The investigation into the complaint was able to confirm the reported event based on the provided photographs.Continuous improvement is of the utmost importance to fresenius medical care as we strive to provide dialysis products of the highest quality to our patients.Reports of leaking product are investigated both individually as complaints, as well as via the nc/capa program, in order to assess and improve our products and processes.Capas for vision systems and blood leak reduction are recent examples of leak related investigations directed at an overall reduction in dialyzer leaks.
|
| |
|