• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ERIKA DE REYNOSA, S.A. DE C.V. 2008T HD SYS. W/O CDX W/BIBAG BLUE STAR; Dialyzer, High Permeability With Or Without Sealed Dialysate System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ERIKA DE REYNOSA, S.A. DE C.V. 2008T HD SYS. W/O CDX W/BIBAG BLUE STAR; Dialyzer, High Permeability With Or Without Sealed Dialysate System Back to Search Results
Catalog Number 191130
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/09/2026
Type of Reportable Event Malfunction
Event or Problem Description
A user facility's biomedical technician (bmt) reported that a 2008t hemodialysis (hd) machine¿s heparin pump was not working and arcing was observed on the distribution board.A fme field service technician (fst) was called onsite and confirmed the arcing as well as discoloration on the heating block and signs that the wires had loosened causing a buildup of heat.The fst replaced the distribution board to resolve the issues.Machine functional tests were completed and passed onsite after repair.It is unknown if there was patient involvement, blood loss, patient serious injury or adverse event.The complaint device is not available to be returned to the manufacturer for evaluation.Multiple attempts have been made to contact the customer to obtain additional information related to the reported event.At this time, no additional information has been provided.
 
Event or Problem Description
A user facility's biomedical technician (bmt) reported that a 2008t hemodialysis (hd) machine¿s heparin pump was not working.During repair the bmt observed arcing and discoloration marks on the 8-pin component of the distribution board.The bmt reported that there was no melting, burning smell, smoke, spark, or flame observed.The machine has not had any past problems with failing the electrical leakage test and there was no damage observed to any other components.The machine has approximately 2,200 hours and is plugged into a hospital grade ground-fault circuit interrupter (gfci) outlet.A fme field service technician (fst) was called onsite and confirmed the arcing as well as discoloration on the heating block and signs that the wires had loosened causing a buildup of heat.The fst replaced the distribution board to resolve the issues.Machine functional tests were completed and passed onsite after repair.There was blood loss, patient serious injury or adverse events reported.The complaint device is not available to be returned to the manufacturer for evaluation.
 
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Additional Manufacturer Narrative
Additional information: b5.The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008T HD SYS. W/O CDX W/BIBAG BLUE STAR
Common Device Name
Dialyzer, High Permeability With Or Without Sealed Dialysate System
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
director, quality systems
900 w zaragosa drive suite d
pharr TX 78577
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
+1(617)417-5172
MDR Report Key50789341
Report Number0008030665-2026-02477
Device Sequence Number10510373
Product Code KDI
UDI-Device Identifier00840861102112
UDI-Public00840861102112
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K243237
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 09/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number191130
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/10/2026
Supplement Date Received by Manufacturer09/18/2026
Initial Report FDA Received Date09/16/2026
Supplement Report FDA Received Date09/18/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/02/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-