| Model Number |
CF-HQ1100DL |
| Medical Device Problem Codes |
Corroded (1131); No Display/Image (1183); Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported the colonovideoscope no image during a inspection for use.During the initial evaluation of the device it was observed, scope communication error b30 and printed circuit board and charged coupled device due to heavy fluid that caused corrosion.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection and the reported failure was confirmed.A root cause could not be identified.Based on the results of the investigation.The most probable cause of the complaint is cause traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|