• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC Ion; System Cart

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC Ion; System Cart Back to Search Results
Model Number 380748-65
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pneumothorax (2012)
Date of Event 08/20/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that after an ion transbronchial lung biopsy procedure, the patient had developed a pneumothorax that required a chest tube.Two nodules were being biopsied and the lesions were 1.7cm located in the right middle lobe (attached to the pleura) and 1.1cm located in the right lower lobe.After biopsying the first nodule, an intraoperative ct scan identified a large pneumothorax, and a 14 gauge chest tube was inserted.The procedure was completed as planned.The physician believed the pneumothorax was caused by a spatial orientation issue and that it could have occurred with another biopsy modality.There were no issues with the ion system.After the procedure, the patient was hospitalized for 2 days then discharged home.The final diagnosis was non-small cell lung cancer.
 
Additional Manufacturer Narrative
There is no report that a malfunction of an ion system, instrument, or accessory occurred.Therefore, no products are expected for return for failure analysis evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Ion
Common Device Name
System Cart
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 95051
+1(408)523-2100
MDR Report Key50789744
Report Number2955842-2026-37991
Device Sequence Number18534394
Product Code EOQ
UDI-Device Identifier00886874116234
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K182188
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 09/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number380748-65
Device Catalogue NumberN/A
Device Lot NumberUnknown
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 08/20/2026
Initial Report FDA Received Date09/16/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/02/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Ion Endoluminal System.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age62 YR
Patient SexFemale
Patient Weight94 LB
-
-