KARL STORZ IMAGING ASM,HEAD,IMAGE1 S,H3-Z FI,CE Back to Search Results
Model Number TH102
Medical Device Problem Code Poor Quality Image (1408)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there was a pixilated haze ring around the edge of the viewing area.The issue could not be solved by changing the lens, but by changing the camera.No negative impact in state of health reported.
 
Additional Manufacturer Narrative
The device in question was returned to the manufacturer.This information is reflected in section d9.The evaluation is anticipated but not yet begun.The investigation will be performed by a designated karl storz employee.The event is filed under internal karl storz complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASM,HEAD,IMAGE1 S,H3-Z FI,CE
Common Device Name
ASM,HEAD,IMAGE1 S,H3-Z FI,CE
Manufacturer (Section D)
KARL STORZ IMAGING
1 s. los cameros road
goleta, ca 93117
GM  93117
Manufacturer (Section G)
KARL STORZ IMAGING
1 s. los cameros road
goleta, ca 93117
GM   93117
Manufacturer Contact
theresa mussaw
2151 e grand ave
el segundo, CA 90245
+1(410)441-1327
MDR Report Key50789752
Report Number2027009-2026-02177
Device Sequence Number18522787
Product Code OWN
UDI-Device Identifier04048551333150
UDI-Public04048551333150
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K152583
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTH102
Device Catalogue NumberTH102
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/03/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/24/2026
Initial Report FDA Received Date09/16/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-