Medline Industries, LP MEDLINE; KIT, LOW BED F/ FULL ELECTRIC HMC BED Back to Search Results
Catalog Number MDSLOWBED
Medical Device Problem Codes Noise, Audible (3273); Damaged Thread (4075)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/26/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the headboard section make noise upon raising and lower.Upon multiple attempts the headboard section would not move any more.
 
Additional Manufacturer Narrative
The customer reported that the headboard section made noise upon raising and lowering and, upon multiple attempts, the headboard section would not move any more.Sample and/or photo/video was not available; medline quality reviewed the documentation listed in the complaint, performed account history/trending and finished good/component trending, and reviewed pcn history and quality control inspection records.The investigation results were not confirmed, and the cause was documented as sample required to determine cause.No trends were identified, no product changes were identified that could have contributed to the alleged product failure, and the division will monitor complaints through quarterly trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
KIT, LOW BED F/ FULL ELECTRIC HMC BED
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
madelyn ruppert
three lakes drive
northfield, IL 60093--275
MDR Report Key50789883
Report Number1417592-2026-42309
Device Sequence Number16094717
Product Code FNL
UDI-Device Identifier10080196361090
UDI-Public10080196361090
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMDSLOWBED
Device Lot Number10824070022
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/26/2025
Initial Report FDA Received Date09/16/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-