| Model Number |
IntelliVue Patient Monitor MX550 |
| Medical Device Problem Code |
Imprecise or Erratic Results (4059)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1: reporting institution (b)(6)
a philips field service engineer (fse) evaluated the device and could not replicate the reported issue.The fse performed accuracy, leakage, and pressure tests using a manual pressure gauge, as well as calibration and accuracy testing per the service manual.All test results were within specified limits, and the system was found to be functioning properly.The device met specifications for the performed service and was returned to use.The device was confirmed to be operating per specifications and no failure was identified.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
|
| |
|
Event or Problem Description
|
|
Philips received a complaint on the intellivue mx550 patient monitor indicating there a variation with the nibp.The device was in use at the time of the event.No adverse event occurred.
|
| |
|