Model Number 60005
Medical Device Problem Code Optical Problem (3001)
Health Effect - Clinical Code Unspecified Eye / Vision Problem (4471)
Date of Event 08/19/2026
Type of Reportable Event Serious Injury
Event or Problem Description
A site reported to rxsight that after a patient's light adjustable lens (lal) was implanted, the patient underwent 3 light adjustments and 1 lock-in treatment postoperatively.The patient returned to the clinic approximately 3 years after implantation and a slit lamp exam found evidence of a suspected central distortion that is consistent with photopolymerization.A fourth light adjustment and second lock-in were then performed.The lal was subsequently explanted and a vitrectomy was performed before a new lal was implanted as a replacement.
 
Additional Manufacturer Narrative
Manufacturing records were reviewed and found no deviations, non-conformances, or other manufacturing issues related to the reported event.Manufacturer reference #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Light Adjustable Lens (LAL)
Common Device Name
Light Adjustable Lens (LAL)
Manufacturer (Section D)
RxSight, Inc.
100 columbia
aliso viejo CA 92656
Manufacturer (Section G)
RxSight, Inc.
100 columbia
aliso viejo CA 92656
Manufacturer Contact
jim zhang
100 columbia
aliso viejo, CA 92656
+1(949)273-6163
MDR Report Key50790249
Report Number3012712027-2026-00397
Device Sequence Number21224974
Product Code PZK
UDI-Device Identifier00818806020654
UDI-Public(01)00818806020654(11)221017(17)250930
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P160055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2025
Device Model Number60005
Device Catalogue Number60005
Device Lot NumberL07-004324
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/19/2026
Initial Report FDA Received Date09/16/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/17/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-