| Model Number |
630-2271211-01 |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
Cardiac Arrest (1762)
|
| Date of Event |
08/22/2026
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
Complainant alleged that while attempting to defibrillate a patient (age & gender unknown) in cardiac arrest, the device was unable to obtain an ecg signal via electrode pads.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that the patient subsequently expired, however it was not a result of the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.Event log review confirmed that ecg was initially obtained through the pads; however, the user switched to lead i rather than pads view.When the analyze button was selected near the end of the event, the device returned to pads view and a clear ecg signal was observed.Device evaluation found no malfunction; the unit passed all functional testing and produced a clear ecg signal using test accessories.The device performed as expected and designed.The reported leads and pads were not returned for evaluation.Based on log review and testing, the reported no ecg condition was attributed to wrong lead view.Analysis of reports of this type has not identified an increase in trend.
|
| |
|